Cone beam CT와 일반 CT의 흡수선량 및 유효선량 비교평가

Comparison of cone beam CT and conventional CT in absorbed and effective dose

초록/요약

목적 : 이 연구는 한국에서 최근 개발된 Cone beam CT와 일반 CT와의 흡수선량및 유효선량을 비교 평가하고자 한다. 재료 및 방법 : 열발광선량계(TLD rod: GR-200 소자, Thermo Fisher ScientificInc., Waltham, MA, USA)를 Male ART Head and Neck Phantom(RadiologySupport Devices Inc., Long Beach, USA)의 25군데에 위치시켰고, 국산 conebeam CT 장비인 Implagraphy(Vatech Co., Hwasung, Korea)와 DCT Pro(VatechCo., Hwasung, Korea), 일반 CT 장비인 SCT-6800TXL(Shimadzu Corp., Kyoto,Japan), 파노라마방사선장비인 Cranex 3+ (Soredex Orion Corp., Helsinki, Finland)를 이용하여 방사선을 조사하였다. Harshaw 3500TLD reader(Thermo FisherScientific Inc., Waltham, MA, USA)로 흡수선량을 측정하였고, 방사선가중선량과유효선량은 2005년 draft ICRP tissue weighting factors를 이용하여 얻었다. 결과 : 흡수선량은 우측악하선부에서 SCT 6800TXL (Implant) 촬영시 110.57mGy로 가장 높았으며, SCT 6800TXL (3D)는 24.56 mGy, Implagraphy3는 22.39mGy, DCT Pro는 7.19 mGy, Implagraphy1는 5.96 mGy, Cranex 3+는 0.70 mGy순이었다. 유효선량 (E2005draft)은 SCT 6800TXL (Implant), SCT 6800TXL (3D), Implagraphy3, DCT Pro, Implagraphy1 순으로 각각 2.551 mSv, 1.272 mSv,0.598 mSv, 0.428 mSv, 0.146 mSv 였고, Cranex 3+의 유효선량 (0.023 mSv)을기준으로 각각 108.6배, 54.1배, 25.5배, 18.2배, 6.2배 였다. 결론 : 국산 cone beam CT장비에서 얻어진 흡수선량과 유효선량은 일반 CT 장비에서 얻어진 흡수선량과 유효선량보다 낮았으며, 상대적으로 낮은 장비가격과개발되어지는 영상모듈로 사용빈도는 계속 증가할 것이다.

more

초록/요약

Purpose : This study provides comparative measurements of absorbed dose and effective dose for newly developed cone beam CT in comparison with them for conventional CT. Materials and Methods : Thermoluminescent dosimeter rods (TLD rod: GR-200 , Thermo Fisher Scientific Inc., Waltham, MA, USA) were placed at 25 sites throughout the layers of Male ART Head and Neck Phantom (Radiology Support Devices Inc., Long Beach, USA) for dosimetry. Implagraphy, DCT Pro(Vatech Co., Hwasung, Korea) units, SCT-6800TXL (Shimadzu Corp., Kyoto, Japan) and Cranex 3+ (Soredex Orion Corp., Helsinki, Finland) were used for radiation exposures. Absorption doses were measured using Harshaw 3500TLD reader(Thermo Fisher Scientific Inc., Waltham, MA, USA). Radiation weighted doses and effective doses were measured and calculated using 2005 draft ICRP tissue weighting factors. Results : Absorbed doses in Rt. submandibular gland were 110.57 mGy for SCT 6800TXL (Implant, 24.56 mGy for SCT 6800TXL (3D), 22.39 mGy for Implagraphy3, 7.19 mGy for DCT Pro, 5.96 mGy for Implagraphy1, 0.70 mGy for Cranex 3+. Effective doses (E2 005draft) were 2.551 mSv for SCT 6800TXL (Implant), 1.272 mSv for SCT 6800TXL (3D), 0.598 mSv for Implagraphy3, 0.428 mSv for DCT Pro and 0.146 mSv for Implagraphy1. These are 108.6, 54.1, 25.5, 18.2 and 6.2 times greater than panoramic examination (Cranex 3+) doses (0.023mSv). Conclusion : Cone beam CT machines recently developed in Korea, showed lower effective doses than conventional CT. Cone beam CT provides a lower dose and cost alternative to conventional CT, promising to revolutionize the practice of oral and maxillofacial radiology.

more

목차

Ⅰ. 서론 = 1
Ⅱ. 연구재료 및 방법 = 3
1. 연구재료 = 3
2. 연구방법 = 3
Ⅲ. 연구결과 = 11
Ⅳ. 총괄 및 고찰 = 15
Ⅴ. 결론 = 18
참고문헌 = 19
영문초록 = 23

more