상용화된 치과용 임플란트의 뒤틀림 제거력 및 조직학적 분석 연구 : 가토 경골에서의 연구
Removal torque and histomorphometric investigation of surface modified commercial implants : An experimental study in the rabbit tibia
- 주제(키워드) histomorphometric evaluation , interfacial shear strength , Micro-CT , Resonance Frequency Analysis , surface modified commercial implant
- 발행기관 江陵大學校 大學院
- 지도교수 Cho, Lee-Ra ; Park, Chan-Jin
- 발행년도 2008
- 학위수여년월 2008. 2
- 학위명 석사
- 학과 및 전공 齒醫學科
- 세부전공 齒科補綴學 專攻
- 원문페이지 34 p.
- 본문언어 한국어
초록/요약
The methods of surface modification of commercial implants were various according to the manufacturer. Surface modification of implant may produce diverse physical and chemical surface characteristics resulted from the treatment method and treatment condition. As a result, the bone response might be different. Even though surface modified implants have been used clinically, most researches are focusing on the bone response of surface modified implants comparing to machined implants rather than surface modified commercial implants. This study compare and analyze bone responses of 4 surface modified commercial implants with different shapes and surfaces. Material and methods Eighty surface modified commercial implants with 4 different surface characteristics were installed in the tibia of white Newzealand rabbits. Biomechanical stability tests and histomorphometric evaluation were done. Results and Conclusions The results were as follows: 1) Surface modified commercial implants showed stable osseointegration at 6 weeks after installation. 2) Histomorphometric evaluation showed that there was no significant differences in bone to implant contact among 4 different commercial titanium implants. In comparing the implants with different shape the measurement of bone growth in subcortical area would be more reliable than entire bone to implant contact length. 3) Resonance Frequency Analysis showed that there was no significant differences among 4 types of implants, even though they were significantly different in installation. 4) There was significant differences in interfacial shear strength among 4 type of implants. 5) It is difficult to observe accurate bone to implant interface using Micro-CT. However, it is possible to measure the entire contact length of the implant to the bone.
more목차
I. 서론 = 1
II. 연구재료 및 방법 = 4
III. 연구 결과 = 9
IV. 총괄 및 고안 = 12
V. 결론 = 15
참고 문헌 = 16
영문 초록 = 22
FIGURES AND TABLES = 24